Instruction Guide – Massa M3 Wireless Sensor Evaluation Kit Page of 14
M3 Kit Guide 111017.docx 10-20-2011
4
PRELIMINARY
M3 Application Software
Validating the Tanks Ultrasonic Performance
(Continued)
5) Mount the M3 Sensor onto the tank.
6) From the “Waveform” tab, select “10 Transmit Cycles and high Gain”. Then click “Acquire
Waveform” to view the ultrasonic signal, as shown in the figure below. It takes a few minutes
to collect all the data from the Sensor. For best operation, close all other Windows
applications.
7) To save the ultrasonic waveforms and status information that were obtained using the “Data”
tab, select “File” and then “Save Sensor to File” from the drop down menu items. It is a good
practice for any new installation to save each waveform indicated in Table 1, especially when
the tank is empty, since this will show if any obstacles exist in the ultrasonic path that may
interfere in obtaining accurate measurements (see the next page for more details). These files
can be recalled for review using the “File” then “Recall Sensor from File” drop down menu
items.
TABLE 1
Some Suggested Waveforms to Save
Waveforms
#1 #2 #3 #4
Sensor Model
Transmit
Cycles Gain
Transmit
Cycles Gain
Transmit
Cycles Gain
Transmit
Cycles Gain
M3/150 and M3/150is 1 low 10 High - - - -
M3/95 and M3/95is 1 low 10 low 10 high 20 high
M3/50 4 low 1 low 10 low 10 high